Medline Industries, Lp: Cardiovascular Procedure Kit
Recall Summary Medline is recalling cardiovascular procedure kits used in open heart and cardiac surgeries because some samples failed quality tests. The problems found include surface irregularities on catheters, connectors that detached prematurely, and catheters with dimensions that didn't meet specifications. The kits were distributed nationwide across the U.S. and internationally to multiple countries.
- Recalling firm
- Medline Industries, LP
- Product
- MEDLINE Medical Procedure Kits labeled as CH OPEN HEART, ADULT CARDIAC KIT, AHT CVOR CABG, APH EMERGENCY OPEN HEART PACK, Etc. (see recall documents for a full list of products)
- Reason
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-05-04
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.