Paragon 28, Inc.: Bone Wedge

Recall Summary Paragon 28, Inc. is recalling the Navicular Cage Small Right Rev A (Model 6123S R), a foot implant device, because the material contains elevated oxygen levels above allowable limits that may reduce its strength and flexibility. The device was distributed nationwide across all U.S. states and Washington, D.C. This is a Class II recall.

Recalling firm
Paragon 28, Inc.
Product
Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant
Reason
Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-07-10
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution i the state of MT.