Howmedica Osteonics Corporation - Trident 0 Degree Constrained Acetabular Insert

Recalling firm
Stryker New Zealand Ltd
Product
Howmedica Osteonics Corporation - Trident 0 Degree Constrained Acetabular Insert. [Device]. Model: Multiple affected lots; Multiple affected lots; Multiple affected lots. Also covers: Constrained Liner; Howmedica Osteonics Corporation - UHR Universal Head Bipolar Component
Reason
The manufacturer has become aware that the devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five (5) years., Industry guidance for storage duration of UHMWPE raw material used in the manufacture of implant bearing/articulating material of prosthetic joints is less than five (5) years. Product manufactured using UHMWPE raw material beyond the recommended storage duration has the potential for elevated levels of oxidation. Oxidation of UHMWPE material can impact material properties, potentially resulting in increased wear.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/316f3c51-806d-4cba-90bb-143c0caa337b