Abiomed, Inc.: Introducer, Catheter

Recall Summary Abiomed is recalling Impella CP Introducer Kits (14 Fr, 13 cm) used with heart pump devices due to a manufacturing defect that can cause fluid leakage from the introducer sheath. The leakage cannot be seen by users and affects introducers included in every pump set. The recall involves products distributed worldwide, including across the United States, Canada, Europe, Asia, and the Middle East.

Recalling firm
Abiomed, Inc.
Product
ABIOMED Impella CP Introducer Kit 14 Fr, 13 cm, Pump Set Code 0052-0038, containing Kit, 14Fr Introducer, 13cm, Long Taper, Introducer Product Code 0052-3052. heart pump and accessories
Reason
A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-11
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT),...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/30ff343a-4d15-41a0-bd39-9278d41267bb