Automated Impella Controller With Impella Connect

Recalling firm
Bio-Excel (NZ) Limited trading as Bio-Excel (Australia) Pty Ltd
Product
Automated Impella controller with Impella Connect. [Device]. Model: Product code: 0042-0040-AU - Optical, AIC, Impella Connect, Pkgd, AU. Affected: All lots and serial numbers
Reason
The manufacturer has identified an increase in complaints due to purge cassette recognition issues resulting from failures of the purge flag component within the purge pressure sensor assembly. The purge flag engages with the purge disc and allows the AIC to recognize that the purge disc is inserted properly. The manufacturer is addressing these failures by replacing the purge flag and purge retainer components within the purge pressure sensor assembly of the AIC.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-21
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE