Angiographic X-ray

Recalling firm
Philips New Zealand Commercial Limited
Product
Angiographic X-Ray. [Device]. Model: Azurion 7 M20; Azurion 7 M20. Affected: System code: 722079; System code: 722224. Also covers: [MDR] Azurion
Reason
The manufacturer identified that, during certain repair or service activities on Azurion systems, SSDs not specified in the Medical Device File may have been installed. These SSDs may have different performance characteristics and potentially reduced reliability over time. If an affected SSD fails to perform as intended, this may result in loss of imaging or motorised movement functionality, a blank/black display, inability to store newly acquired images, or inability to access previously acquired images.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-10-28
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory