[mdr] Pagewriter Tc35 Cardiograph; Cw 5000; Cw7000

Recalling firm
Philips New Zealand Commercial Limited
Product
[MDR] PageWriter TC35 Cardiograph; CW 5000; CW7000. [Device]. Model: Cardiac Workstation 5000. Affected: Product Number - 860439
Reason
The manufacturer received multiple complaints that the Cardiac Workstation may intermittently display an inaccurate, often elevated, native heart rate. While clinicians were generally able to identify the discrepancy because it did not match the patient's clinical presentation, the issue can affect the accuracy of the entire ECG report, including waveform data, measurements, and interpretive statements. The problem is linked to the Wired PIM accessory and may occur, in rare cases, when the Cardiac Workstation exits Standby mode and fails to correctly set the ECG sampling rate. This can...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-07-27
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE