Atellica Ch Revised C-reactive Protein (rcrp)

Recalling firm
Siemens Healthcare (NZ) Ltd
Product
Atellica CH Revised C-Reactive Protein (RCRP). [Device]. Affected: All Lots
Reason
The manufacturer has confirmed through the investigation of customer complaints the potential for falsely elevated results when the Atellica CH Revised C-Reactive Protein (RCRP) assay is used on the Atellica CH and Atellica CI analyzers. A small subset of samples may reproducibly generate one of the following outcomes:, 1. A falsely elevated, reportable result., 2. An initial falsely elevated result above the measuring interval (>25.00 mg/dL or >250.0 mg/L), followed by a 5X auto-diluted result that is inconsistent with the initial result. In this scenario, the auto-diluted result is...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-08-28
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/2a19b8fb-de5c-47b6-b9f3-1f98275c8903