Shockpulse Lithotripsy System

Recalling firm
Olympus (NZ) Ltd
Product
ShockPulse Lithotripsy System. [Device]. Model: ShockPulse-SE Lithotripsy System – Reusable Probes and ShockPulse Lithotripsy Generator. Affected: Serial number - CG6069
Reason
The manufacturer has identified through an internal investigation that specific serial numbers of ShockPulse generators may have a mis-wired component. This condition does not impact the intended functionality of the device; however, the miswiring may introduce additional electrical noise on the power supply output. The presence of noise on the ultrasonic input circuit is not compliant with applicable electromagnetic compatibility (EMC) standards and may adversely affect the overall reliability of the system.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-05
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/29415478-8c73-4041-aff5-c89d73a107ab