Merlin Patient Care System Model 3650
Merlin Patient Care System Model 3650 Recall St Jude Medical is recalling the Merlin Patient Care System Model 3650 due to a software issue that requires an upgrade. This Class I recall affects hospitals, pharmacies, and laboratories in New Zealand that use this system. The recall was issued because the software needs to be updated to address the identified problem.
- Recalling firm
- Abbott Medical New Zealand Limited
- Product
- Merlin Patient Care System Model 3650. [Device]. Model: MERLIN 3650 PATIENT CARE SYSTEM (PCS). Affected: SOFTWARE MODEL 3330 V25.0.X – V28.7.X WHEN USED WITH AVEIR LEADLESS SYSTEM
- Reason
- The manufacturer is contacting customers to inform them of a Merlin PCS 3650 programmer software update that addresses a potential software behaviour observed during the Pacing Capture Threshold (PCT) test of the AVEIR™ Leadless Pacemaker (LP). In rare circumstances, if intermittent loss of telemetry occurs during a PCT test, the programmer may not successfully transmit the command to terminate the test, which could result in the LP remaining in sub-threshold pacing until telemetry is fully disconnected. This issue can only occur during a PCT test, which is performed at the time of implant...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-05-11
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE