Merlin Patient Care System Model 3650

Merlin Patient Care System Model 3650 Recall St Jude Medical is recalling the Merlin Patient Care System Model 3650 due to a software issue that requires an upgrade. This Class I recall affects hospitals, pharmacies, and laboratories in New Zealand that use this system. The recall was issued because the software needs to be updated to address the identified problem.

Recalling firm
Abbott Medical New Zealand Limited
Product
Merlin Patient Care System Model 3650. [Device]. Model: MERLIN 3650 PATIENT CARE SYSTEM (PCS). Affected: SOFTWARE MODEL 3330 V25.0.X – V28.7.X WHEN USED WITH AVEIR LEADLESS SYSTEM
Reason
The manufacturer is contacting customers to inform them of a Merlin PCS 3650 programmer software update that addresses a potential software behaviour observed during the Pacing Capture Threshold (PCT) test of the AVEIR™ Leadless Pacemaker (LP). In rare circumstances, if intermittent loss of telemetry occurs during a PCT test, the programmer may not successfully transmit the command to terminate the test, which could result in the LP remaining in sub-threshold pacing until telemetry is fully disconnected. This issue can only occur during a PCT test, which is performed at the time of implant...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-05-11
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE