Ventilator Circuit For Non-invasive Ventilation

Recalling firm
Active Rehab Equipment
Product
Ventilator circuit for non-invasive ventilation. [Device]. Model: Product Code - AMVC1792-156. Affected: Lot numbers 021025, 181024
Reason
The manufacturer investigation of the customer reports did not identify any anomaly or non-conformance with the product as all devices were manufactured to the specifications set. The existing IFU requires an update to prevent reasonable forseeable misuse when handling the device., Pulling on the tubing may lead to inadvertent tubing separation from the purple cuff connector. Replacement of the circuit may result in delays in treatment and will require intervention by Healthcare Professional. These situations present a potential risk of respiratory distress, hypoxia, and/or low oxygen...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-16
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/278d2bd7-683c-4183-aecb-cf1ab9b9a710