Boston Scientific Corporation: Catheter, Steerable

Recall Summary Boston Scientific is recalling certain FARADRIVE STEERABLE SHEATH CLEAR OUS catheters used to access the heart's blood vessels and chambers because the hemostatic valve can be damaged during device preparation, potentially allowing air to enter the bloodstream. The issue occurs when the dilator is not inserted straight through the center of the valve during setup. This recall affects catheters distributed nationwide and globally.

Recalling firm
Boston Scientific Corporation
Product
FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01
Reason
Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-11
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Global Distribution. US Nationwide.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/2721d258-f407-4956-8a02-1acb771acb4d