I.t.s. Gmbh Autal 28 Lasnitzhohe Austria: Plate, Fixation, Bone

Recalling firm
I.T.S. GmbH Autal 28 Lasnitzhohe Austria
Product
I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301-5, 21301-7, 21301-9, 21301-13, 21302-5, 21302-7, 21302-9, 21302-13. 2. PTL (Proximal Lateral Tibia Locking) Proximal Lateral Tibia Plate; Article Numbers: 21321-7, 21321-12, 21322-7, 21322-12. 3.PTL (Proximal Lateral Tibia Locking) Proximal Medial Tibia Plate; Article Numbers: 21325-4, 21325-7.
Reason
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-01-23
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/27056d12-3f9c-415a-b4f2-8476c3a9cbea