Dxc 500i Clinical Analyzer
- Recalling firm
- Beckman Coulter NZ Ltd
- Product
- DxC 500i Clinical Analyzer. [Device]. Model: DxC 500 AU Module w/ISE, DxC 500i and DxC 500 AU Module, DxC 500i. Affected: Reference: C63522, C63521 - Software: 1.5 or lower
- Reason
- The manufacturer has determined that a modification made by operators to the below 25 predefined calculated rules, which are included as part of the parameter package, will revert to the default expression following parameter package or software upgrades.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-07-16
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/248a9d3c-3657-4afd-a1c5-0872470a3953