Artis Pheno And Artis Icono Systems
Siemens Healthcare is recalling ARTIS pheno and ARTIS icono systems distributed to hospitals, pharmacies, and laboratories in New Zealand due to a software issue that requires an upgrade. This is classified as a Class I recall, which is the most serious type. The recall affects medical imaging systems that were distributed in New Zealand healthcare settings.
- Recalling firm
- Siemens Healthcare (NZ) Ltd
- Product
- ARTIS pheno and ARTIS icono systems. [Device]. Model: Catalogue numbers - 11328100, 10849000, 11327700, 11327600. Affected: ARTIS pheno, ARTIS icono with VE30B and VE40B
- Reason
- The manufacturer has identified that the ARTIS pheno, ARTIS icono with VE30B and VE40B – Incorrectly Reported DAP value. During examination, the system may show sporadically:, 1. that zero dose is administered to the patient. This sporadic behavior is only shown in fluoroscopy and/or acquisitions with one (single) frame. The dose reported is "0", but dose is actually administered to the patient. OR, 2. that a major increase in the dose administered to the patient. The dose reported is significantly higher than the dose actually administered to the patient.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-03-30
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/23918dc8-e602-46ae-9865-49ea67e73b1b