Fresenius Medical Care Holdings, Inc.: Bicarby" Dialysate; Model Numbers: Rfp-400-g, Rfp-401-g, Rfp-402-g, Rfp-404-g, Rfp-407-g, And Rfp-456-g

Recall Summary Fresenius Medical Care is recalling Bicarby Dialysate products (models RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G) due to an increased number of leaks occurring at the luer-lock connection during product preparation. The leaking fluid can create a slip and fall hazard on the floor. The recall affects dialysis patients nationwide who received this product through distribution in New York, Tennessee, and Texas.

Recalling firm
Fresenius Medical Care Holdings, Inc.
Product
Bicarby" Dialysate; Model Numbers: RFP-400-G, RFP-401-G, RFP-402-G, RFP-404-G, RFP-407-G, and RFP-456-G;
Reason
Given the increased number of leaks reported by users, it was determined to be the appropriate course of action to notify users of the potential for the occurrence of Bicarby luer-lock leaks during manipulation of the frangible and tubing when preparing the product for use for treatment. This issue presents the potential of a slip and fall hazard associated with fluid on the floor.
Classification
Class II
Category
medical device recall
Country
United States
Agency
U.S. Food and Drug Administration
Recall date
2026-05-12
Status
Ongoing
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of NY, TN and TX.

Official recall notice from U.S. Food and Drug Administration

Recall details on Nibble Recalls: https://trynibble.app/recalls/2316d5ba-4dbf-49fb-91e4-a2686e8af13e