Ultrasound Systems
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Ultrasound Systems. [Device]. Model: Lumify Systems incompatible with iOS 18 devices
- Reason
- The manufacturer has identified that the Lumify software application has an intentional design component that restricts the ability to scan if the host device is charging. The Lumify Power Module (LPM) provides battery power to support the use of the transducer. It is not intended to provide charging capability to iOS devices. With version 18.X of iOS, when the device's charge level drops below 99%, it will attempt to charge from the LPM through the USB connection. This does not happen for Lightning port connected devices. If the battery threshold is crossed while live scanning, the...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-01-15
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/225fc0b5-88e6-4e44-bd7b-aeedb01ae114