Siemens Medical Solutions Usa, Inc: Interventional Fluoroscopic X-ray System

Recall Summary Siemens Medical Solutions is recalling ARTIS icono ceiling X-ray systems (Model 11328100) because screws in the Display ceiling suspension that holds the monitor in place may gradually loosen over time. This Class II recall affects systems distributed nationwide to healthcare facilities in multiple states. The potential loosening of these screws could cause the monitor to become unstable.

Recalling firm
Siemens Medical Solutions USA, Inc
Product
ARTIS icono ceiling. Model Number: 11328100.
Reason
Potential issue with the Display ceiling suspension (DCS Fix XL) that holds the monitor in place. Over time some screws that keep the monitor attached to the suspension may gradually loosen.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-08
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of CA, GA, IA, IN, LA, MO, NC, NY, PA, SC.