Portable Ventilator, Electric

Recalling firm
ResMed (NZ) Ltd
Product
Portable ventilator, electric. [Device]. Model: Astal 100 150 ventilators
Reason
The manufacturer had been made aware that an issue affecting a subset of Astral 100 and Astral 150 ventilators. An internal electrical component (supercapacitor) may leak electrolyte over time. In rare cases, this may damage specific circuitry on the Ventilator's printed circuit board assembly (PCBA), causing the ventilator to inadvertently enter a fail-safe state
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-25
Status
Ongoing
Distribution
New Zealand - Healthcare Professional

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/207eaddb-c3fa-441b-80b7-b453e7ebc18d