Hysterectomy Kit

Recall Summary Medline's Hysterectomy Pack-LF kits are being recalled nationwide because they contain a defective batch of Mastisol Liquid Adhesive from Ferndale Laboratories. The adhesive's tubing can crack when used, making the product unable to function properly for application. This recall affects medical facilities across all U.S. states and the District of Columbia.

Product
Medline convenience kits: HYSTERECTOMY PACK-LF DYNJ20485L
Reason
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-27
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution. OUS distribution pending.