一般的名称: 超音波手術器 販売名 : ウルトラキュレット
Recall Summary Gunze Medical is recalling 47 UltraCuret ultrasonic surgical handpiece cords (specific serial numbers from August 2025 to July 2026) because the outer tube covering may detach during use. The device is used to remove dead tissue, foreign objects, and infected material from wounds. No serious health problems or injuries have been reported, as the internal wiring remains connected and poses no shock risk even if the outer tube comes off.
- Recalling firm
- 製造販売業者の名称 : グンゼメディカル株式会社 製造販売業者の所在地: 大阪府大阪市北区堂島二丁目4番27号 JRWD堂島タワー 許可の種類 : 第一種医療機器製造販売業 許可番号 : 27B1X00173
- Product
- Lot: 製品名:ハンドピースコード 対象シリアルナンバー 250062,250063,250064,250065,250066,250067,250069,250070,250071,250074,250075, 250083,250084,250092,250095,250096,250097,250098,250099,250100,250101,250102, 250103,250104,250145,250146,250147,250148,250150,250152,250156,250157,250158, 250159,250160,250161,250163,250165,250166,250167,250168,260053,260054,260055, 260056,260057,260059 数量 47個 出荷時期:2025年8月~2026年7月. Use: 本品は、急性創傷、慢性創傷、熱傷等のデブリードマンが必要な患者における死滅又は汚染された組織、異物、 及び感染部位の除去、清浄化のために用いる。. Health hazards: ハンドピースコードのアウターチューブが外れる可能性がありますが、本体及びハンドピースは導線でつながっ ているため継続使用が可能であり、導線の露出による感電の恐れもないため重篤な健康被害に至る不具合が発生...
- Reason
- ハンドピースコードのアウターチューブが外れるという不具合情報から、アウターチューブを固定する部品が組 付けられずに出荷されたことが製造元の調査の結果で確認されましたため、該当するシリアル番号の製品を自主 回収することといたしました。
- Classification
- Class II
- Category
- medical device recall
- Country
- Japan
- Agency
- PMDA-DEVICE
- Recall date
- 2026-10-08
- Status
- Ongoing
Official recall notice from PMDA-DEVICE