Integra Lifesciences Corp. (neurosciences): Retractor

Recall Summary Integra LifeSciences is recalling the OmniTract Surgical Retractor System (633 product codes) because the instructions provided to hospitals and surgical centers state an incorrect drying time of 30 minutes, when the actual required drying time is 90 to 120 minutes depending on the sterilization method used. This product has been distributed worldwide, including throughout the United States and 50+ other countries. This is classified as a Class I recall, the most serious category.

Recalling firm
Integra LifeSciences Corp. (NeuroSciences)
Product
OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.
Reason
The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-05
Status
Open, Classified