Accuray Incorporated: Accelerator, Linear, Medical

Recalling firm
Accuray Incorporated
Product
Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Reason
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-06-16
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of China, Colombia, Czech Republic, France, Hong Kong, Japan, Korea, Republic of, Latvia, Mexico, Myanmar, Saudi Arabia, Taiwan, Turkey, Ukraine.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/1adbd5fc-b28e-43d1-8ae2-762d20a8b6fb