Transpac IV Dual Cable, Dual Ibp Adapter
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Transpac IV Dual Cable, Dual IBP Adapter. [Device]. Model: Dual IBP Adapter. Affected: Product model number: 989803199741
- Reason
- The manufacturer has identified that the Dual IBP Adapter may not provide adequate defibrillation protection when used with non-defibrillation-proof transducers that are not listed as compatible in the IFU. In the event of defibrillation, energy may be partially diverted through the IBP system rather than the patient, potentially creating a hazardous situation. The risk is particularly relevant when catheter-tip transducers are used, as these are typically not defibrillation-proof. The adapter design does not prevent connection to incompatible transducers.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-11-12
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE