Transpac IV Dual Cable, Dual Ibp Adapter

Recalling firm
Philips New Zealand Commercial Limited
Product
Transpac IV Dual Cable, Dual IBP Adapter. [Device]. Model: Dual IBP Adapter. Affected: Product model number: 989803199741
Reason
The manufacturer has identified that the Dual IBP Adapter may not provide adequate defibrillation protection when used with non-defibrillation-proof transducers that are not listed as compatible in the IFU. In the event of defibrillation, energy may be partially diverted through the IBP system rather than the patient, potentially creating a hazardous situation. The risk is particularly relevant when catheter-tip transducers are used, as these are typically not defibrillation-proof. The adapter design does not prevent connection to incompatible transducers.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-11-12
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE