Ge Medical Systems Information Technologies Inc: Monitor, Physiological, Patient(with Arrhythmia Detection Or Alarms)
Recall Summary GE HealthCare is recalling ApexPro CARESCAPE Telemetry Server software versions 5.0 and 6.0 used in patient monitoring systems because they may lose ECG and heart rhythm monitoring data when the central monitoring station loses network connection. The issue affects hospitals and healthcare facilities worldwide that use this telemetry monitoring equipment.
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Product
- GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason
- GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-04-24
- Status
- Open, Classified
- Distribution
- Worldwide distribution.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/1a8dcfb6-3b6d-478e-ae21-c5040c2911e0