Ge Medical Systems Information Technologies Inc: Monitor, Physiological, Patient(with Arrhythmia Detection Or Alarms)

Recall Summary GE HealthCare is recalling ApexPro CARESCAPE Telemetry Server software versions 5.0 and 6.0 used in patient monitoring systems because they may lose ECG and heart rhythm monitoring data when the central monitoring station loses network connection. The issue affects hospitals and healthcare facilities worldwide that use this telemetry monitoring equipment.

Recalling firm
GE Medical Systems Information Technologies Inc
Product
GE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason
GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-24
Status
Open, Classified
Distribution
Worldwide distribution.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/1a8dcfb6-3b6d-478e-ae21-c5040c2911e0