Angiography/angioplasty Kit
Recall Summary Medline is recalling 12 models of angiography and angioplasty kits used in catheterization procedures because equipment calibration problems during sterilization and packaging may have compromised the sterility of these devices. The affected kits were distributed nationwide across all U.S. states and territories. These are medical devices used in hospitals and surgical centers for angiography and angioplasty procedures.
- Product
- Medline Convenience Kits: 1) ANGIO PROCEDURE PACK, Model Number:00-401824AA; 2) ENDO KIT W/SYRINGE, Model Number:DYKE1743; 3) BGMC ANGIOGRAPHY PACK-LF, Model Number:DYNJ0800934D; 4) GENERAL ENDO PACK-LF, Model Number:DYNJ0842873J; 5) ANGIO PROCEDURE PACK, Model Number:DYNJ32328; 6) OR ANGIO PACK-LF, Model Number:DYNJ43415C; 7) ANGIO PACK II, Model Number:DYNJ44066A; 8) SOUTH CATH LAB PACK, Model Number:DYNJ61978A; 9) BIOPSY ANGIO TRAY ALEX, Model Number:DYNJ63542B; 10) SURGICAL TECH KIT, Model Number:DYNJ907452; 11) ANGIOGRAPHY PACK, Model Number:DYNJC2371G; 12) SPECTRUM HEALTH CATH...
- Reason
- Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-01-07
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide - US Nationwide distribution.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/1a68d461-9012-4413-950d-eb2e2ef40750