Evo Iq Lvp
- Recalling firm
- Baxter Healthcare Ltd
- Product
- EVO IQ LVP. [Device]. Model: EVO IQ LVP ANZ - Product code ELVP001ANZ and ELVP002ANZ. Affected: All lots
- Reason
- The manufacturer is issuing a corrective action for EVO IQ Large Volumetric Pumps (LVP) due to a potential issue with occlusion alarms not being triggered. This issue occurs after a downstream occlusion (DSO), or upstream occlusion (USO) initially alarms during an infusion, and the clinician tries to troubleshoot the alarm by opening the door, assessing the IV tubing, and closing the door without resolving the occlusion. This will cause the pump to re-baseline the occlusion pressure and future occlusion alarms may be delayed., 18/03/2025 Update - The manufacturer intends to issue a...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-03-18
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE
Recall details on Nibble Recalls: https://trynibble.app/recalls/195424dc-99ca-4075-ac08-8c530d9aac4e