Evo Iq Lvp

Recalling firm
Baxter Healthcare Ltd
Product
EVO IQ LVP. [Device]. Model: EVO IQ LVP ANZ - Product code ELVP001ANZ and ELVP002ANZ. Affected: All lots
Reason
The manufacturer is issuing a corrective action for EVO IQ Large Volumetric Pumps (LVP) due to a potential issue with occlusion alarms not being triggered. This issue occurs after a downstream occlusion (DSO), or upstream occlusion (USO) initially alarms during an infusion, and the clinician tries to troubleshoot the alarm by opening the door, assessing the IV tubing, and closing the door without resolving the occlusion. This will cause the pump to re-baseline the occlusion pressure and future occlusion alarms may be delayed., 18/03/2025 Update - The manufacturer intends to issue a...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-03-18
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE