Trilogy Evo

Recalling firm
Philips New Zealand Commercial Limited
Product
Trilogy EVO. [Device]. Model: Trilogy Evo, Trilogy Evo O2; Trilogy EV300. Affected: Update to latest software version 1.06.15.00.; Update to latest software version 1.06.15.00.. Also covers: Trilogy EV300
Reason
The manufacturer has become aware that Trilogy Evo, Trilogy Evo O2, and Trilogy EV300 devices do not comply with the Obstruction Alarm requirements specified within ISO 80601-2-12 Clause 201.12.4.108 and ISO 80601-2-72 Clause 201.12.4.107., Philips Respironics has assessed this problem and determined that there is a potential for patients that are using a Trilogy Evo device to experience harm as a result of a delay in the obstruction alarm. The delay in an obstruction alarm could result in a delay of a patient or caregiver addressing the problem causing: Respiratory Discomfort,...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-23
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/19135140-3382-440c-975e-4fd24154b83a