Bd Connecta Stopcock

Recalling firm
Becton Dickinson Ltd
Product
BD Connecta stopcock. [Device]. Model: BD Connecta stopcock with BD Q-Syte luer access split septum, BD Connecta stopcock without extension tube, BD Connecta Stopcock with OFF tap without extension tube, BD Connecta Stopcock with extension tube, and injection value. Affected: All Lot numbers
Reason
The manufacturer identified an issue with Port C of the BD Connecta Stopcocks when used with polypropylene luer lock syringes. In some cases, the connection may over-thread and continue to spin, which can be perceived by users as unstable., Testing confirmed that the connection remains secure with no leakage under clinically relevant conditions, and the issue does not occur when used with polycarbonate syringes or IV tubing. The behaviour is due to lower torque required to override the threading when polypropylene syringes are used., While no functional failure was identified, this may...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-18
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE