Philips Mr Systems With Sw Versions R5 & R11 - Ingenia, Achieva Series
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Philips MR systems with SW versions R5 & R11 - Ingenia, Achieva Series. [Device]. Model: Ingenia Elition X, Ingenia Elition S, Ingenia 1.5T, SmartPath to dStream for 1.5T. Affected: Model numbers - 781358, 781357, 781341, 781260
- Reason
- The manufacturer has become aware of a software problem which may terminate an examination during the image reconstruction process. If the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory, the system will freeze, preventing image generation and requiring a system re-start., This problem can occur when the 3D Vane scan protocol is customised with a reconstruction matrix greater than 480 and/or more than 167 slices. Due to a system error in memory budgeting, the system may allow a scan to start even when there is not enough...
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-09-14
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE