Reason
Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning...