Philips Medical Systems Nederland B.v.: Interventional Fluoroscopic X-ray System

Philips is recalling its Azurion 5 M12 interventional X-ray system because some units may contain hard drives not approved for the device, which could cause imaging problems or system failures like loss of image storage, blank screens, or inability to move motorized components. The recall affects these systems distributed worldwide, including across all US states. This is a Class II recall, indicating a serious issue that could result in adverse health consequences but is unlikely to cause death.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion 5 M12. Model Number: 722227.
Reason
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-10-06
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/1673491f-4a65-4954-86bb-f599d1606259