Cesarean Section Tray

Recall Summary Medline is recalling four cesarean section surgical kits that contain a defective adhesive product made by Ferndale Laboratories. The tubing in the adhesive applicator may crack when used, making the adhesive unusable during surgery. The affected kits have been distributed nationwide across all U.S. states and territories.

Product
Medline convenience kits: C SECTION CDS985505G C SECTION PACK DYNJ68260D C-SECTION DYNJ905094B C-SECTION RECOVERY #2 CDS820091P
Reason
Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-27
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution. OUS distribution pending.