Oculus Technologies Of Mexico, S.a. De C.v. Calle Industria Vidriera 81 Industrial Zapopan Norte Zapopan Mexico: Dressing, Wound, Drug

PhaseOne Antimicrobial Solution Recall Summary Oculus Technologies of Mexico is recalling PhaseOne Antimicrobial Solution (Model 15050) because the stainless steel spring in the sprayer component is corroding due to incompatibility with the solution. This corrosion can cause visible discoloration and release trace amounts of metallic elements like iron, chromium, and nickel into the product during use. The product was distributed nationwide across most U.S. states.

Recalling firm
Oculus Technologies of Mexico, S.A. de C.V. Calle Industria Vidriera 81 Industrial Zapopan Norte Zapopan Mexico
Product
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
Reason
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-27
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/163c72f2-4e5a-4e1c-91d9-388e35295ad2