Mindray Beneheart C Series Aed

Recalling firm
Connected Healthcare Systems
Product
Mindray BeneHeart C series AED. [Device]. Model: BeneHeart C-Series Automated External Defibrillator
Reason
The operator's manual is being updated to help users better understand the impact of the Wi-Fi/cellular settings on battery life. Details of the modifications are as follows:, When the equipment is configured with the Wi-Fi/cellular module, it automatically sends self-test reports at preset intervals by default, which consumes battery power. Information has been updated in the manual to remind users that incorrect configuration of the wireless network would cause extra power consumption. Low battery power may cause the device to fail to power on, delaying therapy.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-20
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/14ece249-d1f1-4857-b226-c0f300cd26af