Philips Heartstream Efficia Dfm100 Defibrillator/monitor

Recalling firm
Clinical & Regulatory Services Limited
Product
Philips HEARTSTREAM Efficia DFM100 Defibrillator/Monitor. [Device]. Model: HEARTSTREAM Efficia DFM100 Defibrillator/Monitor. Affected: Efficia DFM100 devices with software revision 2.00.33 installed
Reason
The Efficia DFM100 Defibrillator/Monitor, with software revision 2.00.33 installed, may fail to complete its boot-up sequence. This could occur either 1) during manual boot-up if the knob is turned to AED/Manual/Pacer or 2) as part of a self-test if the device is turned off and connected to any power source. Instead of completing the boot-up sequence, the device may repeatedly display the splash screen and will not be available for clinical use. This will be indicated to the user by displaying a solid red X in the Ready For Use (RFU) indicator.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-08
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE