The Venous Bubble Sensor

Recalling firm
Getinge Australia Pty Ltd (NZ)
Product
The Venous Bubble Sensor. [Device]. Model: BS 3/8x3/32 L1.7 Mat. 701055720 (Bubble Sensor for 3/8 x 3/32 tubing, length 1,7 m)
Reason
The manufacturer has identified a problem with the durability of the connecting cable near the connection to the Bubble Sensor. Excessive bending of the cable can lead to poor contact, which may trigger the errors Ven. bubble sensor defective or Ven. bubble sensor disconnected on the connected medical device (ROTAFLOW II or CARDIOHELP-i). These errors can occur temporarily when the cable is moved or permanently if the connection is fully compromised.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-20
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE