Philips Medical Systems Nederland B.v.: Philips Azurion S. Model Number: 722285
Recall Summary Philips is recalling Azurion S imaging devices (Model 722285) worldwide because some units may contain hard drives that weren't approved for use in these medical devices. These unapproved drives may fail over time, potentially causing the imaging system to lose functionality, display problems, or inability to save or access images. The recall affects devices distributed across the United States and numerous countries globally.
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion S. Model Number: 722285.
- Reason
- Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- U.S. Food and Drug Administration
- Recall date
- 2026-10-06
- Status
- Ongoing
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...