Philips Mr Systems Ingenia, Achieva Series

Recalling firm
Philips New Zealand Commercial Limited
Product
Philips MR systems Ingenia, Achieva Series. [Device]. Model: Philips MR systems with SW version R11.1 and R12.1
Reason
The manufacturer has identified potential stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture, Archiving and Communication System (PACS)., The MRE viewer PACS indicates around 2 kPa (incorrect stiffness values), which can be associated with healthy tissue, while the general viewing environment in the MR Console viewer indicates almost 8 kPa (correct stiffness values), which can be an indicator of cirrhosis.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-01-14
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE