Philips Medical Systems Nederland B.v.: Interventional Fluoroscopic X-ray System

Recall Summary Philips is recalling certain Azurion 3 M15 X-ray systems because they may contain hard drives not approved for use in these devices, which could cause the system to lose imaging capability, display a blank screen, or fail to store or retrieve images. The recall affects systems distributed worldwide, including throughout the United States. This is a Class II recall, indicating a potential serious health risk.

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Philips Azurion 3 M15. Model Number: 722064, 722222, 722280.
Reason
Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed in various PCs. These SSDs may have different performance characteristics, which may include reduced reliability over time such as loss of imaging functionality or motorized movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-10-06
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussal, Bulgaria, Cambodia, Canada, Cayman Islands,...

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/137e2348-3012-4178-b65b-75a9c1a23df5