Rida®plex Respiratory Pathogens C 16 Well Step 1

Recall Summary AusDiagnostics is recalling RIDA®Plex Respiratory Pathogens C 16 well Step 1 (Lot S3520206), a respiratory disease testing device distributed to hospitals, pharmacies, and laboratories in New Zealand. The device has increased background signal in the RSV-A/B detection channel that may cause incorrect test results, including false positive readings for respiratory syncytial virus. This is a Class I recall, indicating the highest level of concern.

Recalling firm
AusDiagnostics NZ Ltd
Product
RIDA®Plex Respiratory Pathogens C 16 well Step 1. [Device]. Affected: Affected Lot S3520206
Reason
An increase in background signal in the RSV-A/B detection channel for affected Lot S3520206 was identified. This effect may generate atypical RSV-A/B amplification profiles resulting in incorrect RSV-A/B Check or false Positive calls.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-06-30
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE