[mdr] Spectral Ct, Iqon Spectral Ct, Brilliance Ict, Brilliannce Ict Upgrade

Recall Summary Philips is recalling certain CT scanner models (Spectral CT, IQon Spectral CT, Brilliance iCT, and Brilliance iCT Upgrade) with specific software versions due to multiple technical issues. The problems include image quality defects during certain scan types, incorrect measurements in some reconstructed images, unexpected movement of the scanner couch during procedures, and potential rotor rotation hazards during maintenance. This recall affects CT systems used in hospitals and laboratories in New Zealand.

Recalling firm
Philips New Zealand Commercial Limited
Product
[MDR] Spectral CT, IQon Spectral CT, Brilliance iCT, Brilliannce iCT Upgrade. [Device]. Model: Philips Spectral CT with software version V5.0, V5.2, or V5.4; Philips Spectral CT with software version V5.0, V5.2, or V5.4. Affected: Product code: 728333; Product code: 728333. Also covers: BEST CT Systems
Reason
Philips has identified software issues that may affect the safety and performance of certain CT systems:, 1., Image artefacts at 80 kVp: Hyperdense, hypodense, or other image quality artefacts may occur during 80 kVp scans., 2. Image artefacts in off-centre brain scans: Brain scans performed with the head positioned off-centre may exhibit image artefacts in the bone-brain area., 3. Incorrect slice thickness with EFOV: Images reconstructed using an extended field of view (EFOV) may have an actual slice thickness different from the displayed value., 4., Unexpected couch movement: During...
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-09-10
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE

Recall details on Nibble Recalls: https://trynibble.app/recalls/107ce0fc-8f2a-402e-b2f7-2ea7b074eb55