Ventec Life Systems, Inc.: Ventilator, Continuous, Facility Use

Recall Summary Ventec Life Systems is recalling VOCSN+Pro ventilators used in medical facilities because some units may have a defective backup battery that prevents the backup alarm from sounding during a power loss. This could delay caregivers from noticing a hazardous situation. The affected ventilators were distributed across the United States.

Recalling firm
Ventec Life Systems, Inc.
Product
VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001
Reason
Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-08
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the state of UT.