Mobile Digital X Ray System

Recalling firm
Philips New Zealand Commercial Limited
Product
Mobile Digital X Ray System. [Device]. Model: Philips MobileDiagnost wDR
Reason
If excessive liquid is used during cleaning of the MobileDiagnost wDR system, fluid can enter the handgrip button edges and cause corrosion or an electrical short at the overlay connector. This may result in unintended slow movement of the system (20�cm/s), controlled by the four handgrip motion buttons. At this low speed, users or bystanders can easily move away, and the system can be stopped using the Emergency Stop button. The issue resolves once the liquid dries.
Classification
Class I
Category
medical device recall
Country
New Zealand
Agency
MEDSAFE
Recall date
2026-02-17
Status
Ongoing
Distribution
New Zealand - Hospital/Pharmacy/Laboratory

Official recall notice from MEDSAFE