Mobile Digital X Ray System
- Recalling firm
- Philips New Zealand Commercial Limited
- Product
- Mobile Digital X Ray System. [Device]. Model: Philips MobileDiagnost wDR
- Reason
- If excessive liquid is used during cleaning of the MobileDiagnost wDR system, fluid can enter the handgrip button edges and cause corrosion or an electrical short at the overlay connector. This may result in unintended slow movement of the system (20�cm/s), controlled by the four handgrip motion buttons. At this low speed, users or bystanders can easily move away, and the system can be stopped using the Emergency Stop button. The issue resolves once the liquid dries.
- Classification
- Class I
- Category
- medical device recall
- Country
- New Zealand
- Agency
- MEDSAFE
- Recall date
- 2026-02-17
- Status
- Ongoing
- Distribution
- New Zealand - Hospital/Pharmacy/Laboratory
Official recall notice from MEDSAFE