Medline Industries, Lp: General Surgery Tray

Recall Summary Medline Convenience surgical kits containing Swan-Ganz catheters are being recalled because the catheters have a defective blue hub on one of their fluid delivery lines that can leak, break, or damage the lumen. This defect can result in medication leakage, blood loss, or infection. The recalled kits were distributed nationwide, with documented distribution in Arizona, California, Florida, Illinois, Missouri, New Jersey, Ohio, Oklahoma, Tennessee, Texas, and Wisconsin.

Recalling firm
Medline Industries, LP
Product
Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B
Reason
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-05-28
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/0e7f0926-0158-41d6-bb6c-8157fc223c53