Medline Industries, Lp: General Surgery Tray
Recall Summary Medline Convenience surgical kits containing Swan-Ganz catheters are being recalled because the catheters have a defective blue hub on one of their fluid delivery lines that can leak, break, or damage the lumen. This defect can result in medication leakage, blood loss, or infection. The recalled kits were distributed nationwide, with documented distribution in Arizona, California, Florida, Illinois, Missouri, New Jersey, Ohio, Oklahoma, Tennessee, Texas, and Wisconsin.
- Recalling firm
- Medline Industries, LP
- Product
- Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG DYNJ905855 KIT UNIVERSITY CABG PACK DYNJ905857 OH MED RN CDS DYNJ905936B
- Reason
- Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-05-28
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/0e7f0926-0158-41d6-bb6c-8157fc223c53