Generic Name: Active Extension And Flexion Rotation Exercise Device Trade Name: Wellwalk

Recall Summary Well Walk exercise equipment (model WW-3000, serial numbers 550-0003 through 550-0014) is being recalled because one of multiple protective grounding wires in the right robotic leg component may wear through and break after a year of use due to improper contact during assembly. The equipment is used to maintain, develop, or recover lower limb muscle strength, and was shipped between July 8, 2024 and January 22, 2025. The manufacturer has already changed its assembly process to prevent the problem going forward and will voluntarily repair existing units to inspect and correct the wiring.

Recalling firm
Name of Manufacturer/Distributor: Good Life Design Co., Ltd. Address of Manufacturer/Distributor: 1-62 Heiwa-cho, Toyota City, Aichi Prefecture Type of License: Type II Medical Device...
Product
Lot: 対象型式:WW-3000 対象製造番号: 550-0003,550-0004,550-0005,550-0006,550-0007,550-0008,550-0009,550-0010,550-0011,550-0012,550-0013, 550-0014 出荷数量:12台 出荷時期:令和7年7月8日から令和8年1月22日. Use: 下肢の筋力を維持、発達または回復させるために用いる. Health hazards: 当該不具合(アース線の一本が断線)の発生によっても、断線したアース線以外にも複数のアース線を設置して おりますので、ロボット脚駆動用モータ筐体等で漏電が発生した際にも感電する恐れはなく、重篤な健康被害の 発生はありません。 これまでに、本事象に起因する健康被害の報告はありません。
Reason
Depending on the manufacturer’s assembly method during the assembly process for this product, it has been confirmed that one of the protective grounding (hereinafter referred to as “ground”) may come into excessive contact with the structural frame. It has been confirmed that, if used in this condition for over one year, the wiring insulation may wear away, potentially causing the ground wire to break. Although the assembly method has been modified in the manufacturing plant’s assembly process to prevent this defect, we are implementing a voluntary product recall to inspect and repair the...
Classification
Class II
Category
medical device recall
Country
Japan
Agency
PMDA-DEVICE
Recall date
2026-03-11
Status
Ongoing