Beckman Coulter Biomedical Gmbh: Calculator/data Processing Module, For Clinical Use

Recalling firm
Beckman Coulter Biomedical GmbH
Product
REMISOL Advance with Software Version 2.5 UDI-DI code: 15099590802073 The Normand Info REMISOL Advance Data Manager collects and manages data and workflow for 128 connections
Reason
Due to software issues where does not handle combo assay results. The software may replace the final combo test result with one of the sub-results resulting in incorrect reporting of test outcomes.
Classification
Class I
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-04-29
Status
Open, Classified
Areas affected
FL, GA, IL, MI, PA
Distribution
U.S.: FL, GA, IL, MI, and PA O.U.S.: Belgium, France, Hungary, Poland, Taiwan (Province of China), Tunisia, United Arab Emirates, and United Kingdom of Great Britain and Northern Ireland.

Official recall notice from FDA-DEVICES

Recall details on Nibble Recalls: https://trynibble.app/recalls/0cb18c7c-d1dc-4644-9891-f0edc6b83d75