Elekta, Inc.: Neurological Stereotaxic Instrument

Recall Summary Elekta, Inc. is recalling the Leksell Vantage Arc System, a neurological surgical instrument used in procedures, because locking mechanisms may fail to properly secure to the Arc system, which could cause coordinate shifts during surgery. The device has been distributed across multiple U.S. states and numerous countries worldwide. This is classified as a Class II recall.

Recalling firm
Elekta, Inc.
Product
Brand Name: Leksell Vantage Arc System REF: 1053958
Reason
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2026-03-19
Status
Open, Classified
Areas affected
AZ, AR, CA, CO, FL, GA, ID, IL, KS, MD, MA, MO, ...
Distribution
US: AR AZ CA CO DC FL GA ID IL KS MA MD ME MO MT NC NE NH NJ NM NV NY OH OK PA SC TX UT VA WA WI WV OUS: Algeria Australia Austria Azerbaijan Belgium China Colombia Costa Rica Croatia Czech Republic Egypt Estonia Finland France Germany Greece Hong Kong Hungary India Israel Italy Japan Korea,...