Beta Bionics, Inc.: Alternate Controller Enabled Insulin Infusion Pump

Recalling firm
Beta Bionics, Inc.
Product
iLet Bionic Pancreas, REF: BB1001
Reason
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal...
Classification
Class II
Category
medical device recall
Country
United States
Agency
FDA-DEVICES
Recall date
2025-09-23
Status
Open, Classified
Areas affected
AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
Distribution
US: WI, HI, OH, ME, CO, NJ, NE, KY, SC, CA, TX, NH, NY, WA, PA, MD, FL, IA, GA, MI, VT, TN, IL, IN, KS, MA, LA, MS, NC, CT, MN, VA, OK, AR, AZ, MO, MT, RI, DC, NM, OR, ID, NV, AK, ND, DE, WV, WY, SD, UT, AL

Official recall notice from FDA-DEVICES