Baxter Healthcare Corporation: Device, Positive Pressure Breathing, Intermittent
Recall Summary Baxter Healthcare is recalling the Hillrom VOLARA System P.CIRCUIT 5KIT AC, a positive pressure breathing device used for lung expansion therapy. The nebulizer cup can leak air and medication during treatment if not properly locked after medication is added, which may reduce oxygen levels and prevent proper medication delivery to the patient. The recall affects devices distributed nationwide in the US and Canada.
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device
- Reason
- There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
- Classification
- Class II
- Category
- medical device recall
- Country
- United States
- Agency
- FDA-DEVICES
- Recall date
- 2026-05-21
- Status
- Open, Classified
- Areas affected
- AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, ...
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
Official recall notice from FDA-DEVICES
Recall details on Nibble Recalls: https://trynibble.app/recalls/0606cf55-3027-42bc-83e3-aa06fe4d4c17